The newly developed spectrometer in the measurement set-up
With the spectrometer featured here, transmission can be measured angle-dependently in a wavelength range of 240 to 1000 nm with an angle of divergence of 0.25 °. The divergence angle set is reduced by a factor of 10 in comparison with a commercial spectrometer. In the measurements the monochromator can be set to a minimum spectral bandwidth of 0.06 nm. Since the spectrometer's measuring beam must have a maximum intensity at a low divergence angle for the measurements, a xenon arc lamp is used as the light source with this measurement configuration. In contrast to other light sources, the xenon arc lamp has a comparatively small light-emitting area and a high light density. Via a mirror and aperture system, a polychromatic measuring beam with a low divergence is generated. After passing through the interference filter the measuring beam is input into a monochromator with a high-resolution diffraction grating and picked up by a low-noise detector. The measurement set-up is equipped with a chopper and can thus measure optical densities up to 4.
Test measurements with different angles of divergence
In order to show the broadenings of the structures as a function of the divergence angle, a sharp transmission structure of an interference filter was examined more closely. The filter was measured with a spectral bandwidth of 0.5 nm and an angle of incidence of 45 ° at the divergence angles of 0.25 ° – 0.88 ° – 1.15 ° – 1.63 ° – 2.06 ° (see top graph). During examination of the widths of the individual peaks, a quadratic relationship was found between half-width values and divergence angles. The lower graph shows the results from this test. Measuring beams with a small divergence angle were therefore required in order to characterize band-pass filters with small half-widths or long-pass filters with spectrally steep edges.
Outlook
The measuring range of optical density is to be extended in the future in order to resolve even lower intensities. This will enable the characterization, among other things, of narrow notch filters with very high blocking (> OD 7).